You are here: Modules Pulse Oximetry PEARL 100  

PEARL 100




SpO2 Module with Optimized Artefact Suppression

  • Advanced signal processor technology with fourth generation algorithm and artefact suppression.
  • Algorithm: "PEARL"(*), Medlab’s new SpO2 technology.
  • Lock-in based signal recovery down to a pulsation of 0.025 % (AC/DC ratio)
  • Data Processing: In the time domain
  • Small size of 77 x 53 mm, max. height 6 mm, electrically and mechanically compatible with the EG00302.
  • Single power supply, 2.5 - 5.5 V DC, 90 mW (@3.3 V).
  • High flexibility in interface protocols.
  • Measures pulse rates from 30 - 250 bpm.
  • Measures SpO2 from 0 .. 100 % (no accuracy stated for values less than 70 %).
  • Clinical validation data available to support certification and registrations of the final product

* Pulse Enhancement Artefact Rejection Logic



Data Sheet

Download data sheet [155 KB] in PDF format.

Technical Manual

Technical Manual [420 KB] in PDF format



Evaluation Software

Test Software for evaluation of the PEARL100 board. Visual Basic 6.0, with full source code and compiled program. Download [97 KB] and extract into one directory.



Test kit PEARL100
Order Nr: 03005

Contents:
• PEARL100 OEM Board
• Finger Probe
• DSUB Amp Adapter
• PC / Power Cable
• Evaluation software